ASTM E1002-2011(2018) 6250 Standard Practice for Leaks Using Ultrasonics.pdf

上传人:赵齐羽 文档编号:527708 上传时间:2018-12-05 格式:PDF 页数:4 大小:109.21KB
下载 相关 举报
ASTM E1002-2011(2018) 6250 Standard Practice for Leaks Using Ultrasonics.pdf_第1页
第1页 / 共4页
ASTM E1002-2011(2018) 6250 Standard Practice for Leaks Using Ultrasonics.pdf_第2页
第2页 / 共4页
ASTM E1002-2011(2018) 6250 Standard Practice for Leaks Using Ultrasonics.pdf_第3页
第3页 / 共4页
ASTM E1002-2011(2018) 6250 Standard Practice for Leaks Using Ultrasonics.pdf_第4页
第4页 / 共4页
亲,该文档总共4页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: E1002 11 (Reapproved 2018)Standard Practice forLeaks Using Ultrasonics1This standard is issued under the fixed designation E1002; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in p

2、arentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 Practice A, PressurizationThis practice covers proce-dures for calibration of ultrasonic instruments, location, andestimated measurements of gas l

3、eakage to atmosphere by theairborne ultrasonic technique.21.2 In general practice this should be limited to leaksdetected by two classifications of instruments, Class I andClass II. Class I instruments should have a minimum detectableleak rate of 6.7 107mol/s (1.5 102std. cm3/s at 0C) ormore for the

4、 pressure method of gas leakage to atmosphere.Class II instruments should have a minimal detectable leak rateof 6.7 106mol/s (1.5 101std. cm3/s at 0C) or more forthe pressure method of gas leakage to atmosphere. Refer toGuide E432 for additional information.1.3 Practice B, Ultrasonic TransmitterFor

5、object undertest not capable of being pressurized but capable of havingultrasonic tone placed/injected into the test area to act as anultrasonic leak trace source.1.3.1 This practice is limited to leaks producing leakage of6.7106mol/s (1.5 101std. cm3/s at 0C) or greater.1.4 The values stated in SI

6、units are to be regarded as thestandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to consult andestablish appropriate safety, health, and environmental prac-tices and determine the ap

7、plicability of regulatory limitationsprior to use.1.6 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by

8、 the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:3E432 Guide for Selection of a Leak Testing MethodE543 Specification for Agencies Performing NondestructiveTestingE1316 Terminology for Nondestructive Examinations2.2 Other Documents:SN

9、T-TC-1A Recommended Practice for Personnel Qualifi-cation and Certification in Nondestructive Testing4ANSI/ASNT CP-189 ASNT Standard for Qualification andCertification of Nondestructive Testing Personnel4NAS-410 Certification and Qualification of NondestructivePersonnel53. Terminology3.1 Definitions

10、For definitions of terms used in thispractice, see Terminology E1316, Section E.4. Summary of Practice4.1 Practice AThis practice sets minimum requirementsfor an ultrasonic detector (Classifications I and II). It providesfor sensitivity validation of the detector and gives proceduresfor pressurizing

11、 the test object, locating leaks, and estimatingthe leakage rates.4.2 Practice BThis practice sets minimum requirementsfor an ultrasonic detector used in conjunction with an ultra-sonic transmitter. It gives procedures for locating leaks usingan electronically generated ultrasonic leak tracer source

12、.5. Basis of Application5.1 The following items are subject to contractual agree-ment between parties using or referencing this practice.5.2 Personnel Qualification1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommitt

13、ee E07.08 on LeakTesting Method.Current edition approved June 1, 2018. Published July 2018. Originally approvedin 1986. Last previous edition approved in 2011 as E1002 - 11. DOI: 10.1520/E1002-11R18.2This technique is sometimes called “ultrasonic translation.”3For referenced ASTM standards, visit th

14、e ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available fromAmerican Society for Nondestructive Testing (ASNT), P.O. Box28518, 1711 Arlingate Ln.,

15、 Columbus, OH 43228-0518, http:/www.asnt.org.5Available from Aerospace Industries Association of America, Inc. (AIA), 1000Wilson Blvd., Suite 1700,Arlington, VA22209-3928, http:/www.aia-aerospace.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959.

16、United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barr

17、iers to Trade (TBT) Committee.15.2.1 If specified in the contractual agreement, personnelperforming examinations to this practice shall be qualified inaccordance with a national or internationally recognized NDTpersonnel qualification practice or standard such as ANSI/ASNT-CP-189, SNT-TC-1A, NAS-410

18、, or a similar documentand certified by the employer or certifying agency, as appli-cable. The practice or standard used and its applicable revisionshall be identified in the contractual agreement between theusing parties.5.3 Qualification of Nondestructive AgenciesIf specifiedin the contractual agr

19、eement, NDT agencies shall be qualifiedand evaluated as described in Practice E543. The applicablerevision of Practice E543 shall be in the contractual agreement.6. Significance and Use6.1 Practice AThis practice is useful for locating andestimating the size of pressurized gas leaks, either as a qua

20、litycontrol test or as a field inspection procedure. It is also valuableas a pretest before other more time consuming and moresensitive leak tests are employed. It should not be usedexclusively to locate highly toxic or explosive gas leaks.6.2 Practice BThis practice is useful for locating leaks ins

21、ystems that are not under pressure or vacuum as either aquality control or a field inspection procedure. It is not usefulfor estimating the size of a leak. It is also valuable as a pretestbefore leak tests using pressurized gas methods and moresensitive leak tests are employed.7. Interferences7.1 Th

22、e areas to be tested must be free of oil, grease, paint,and other contaminants that might mask a leak.7.2 Under certain conditions background noise detected bythe instrument can prevent the detection of relevant leakage.This background noise can result from equipment vibration andair movement due, f

23、or example, to wind, or air-cooled motors,aircraft engines, pneumatic systems, etc.7.3 Use of earphones is required in areas where the back-ground noise might interfere with hearing the audible output ofa speaker.8. Apparatus8.1 Ultrasonic Leak Detection System:8.1.1 The system shall consist of an i

24、nstrument, probes,focusing probe accessory and earphones. (A speaker may ormay not be utilized.)8.1.2 The system shall provide for detection of acousticenergy in the ultrasonic range from 20 000 to 100 000 Hz andshall translate this energy into an audible signal that can beheard by use of earphones

25、or speaker, or both.8.1.3 The detected energy shall be indicated on a signallevel indicator.8.2 Minimum Instrument RequirementsThe instrumentshall meet the following requirements:8.2.1 The detected ultrasonic energy shall be indicated on asignal level indicator with a change above normal backgroundf

26、luctuations.8.2.2 The audible response shall consist of the down-converted heterodyned ultrasonic signal. This audio signal willbe representative of the amplitude or frequency characteristics,or both, of the original ultrasonic signal. Heterodyned signalscould allow the operator to discriminate audi

27、ble backgroundinterference as in 7.2.8.2.3 The instrument shall be equipped with a sensitivitycontrol or gain adjustment, or both, to achieve the conditions of9.3.3.8.2.4 The internal power supply shall be regulated toprovide repeatability of the sensitivity set point to within 65%of full scale over

28、 the full range of battery condition.8.2.5 The instrument may include additional features whichcould enhance leak detection such as frequency controls ormeter/display response selection, or both.8.3 Other ApparatusPressure gages, valves, and piping asrequired.8.4 Ultrasonic Transmission8.4.1 The sys

29、tem shall consist of an apparatus described in8.1 8.3 with the addition of an ultrasonic transmitter.8.4.2 The transmitter shall produce an ultrasound peakedbetween the frequencies of 38 to 42 kHz.8.4.3 The output of the transmitter transducer shall be inexcess of 100 dB at 1 ft.8.4.4 The intensity

30、of the output may be adjustable.8.4.5 The heterodyned signal received by the ultrasonic leakdetection system from the ultrasonic transmitter must be easilyidentified and recognized as the ultrasonic tracer source.8.4.6 The internal power supply shall be regulated and besufficient to produce the requ

31、ired output (see 8.4.3).8.4.7 The transmitter may include additional features thatcould enhance the leak detection process such as a warble tonetransmission and amplitude adjustments.9. Calibration9.1 Calibration/Sensitivity ValidationThe ultrasonic in-strument should be calibrated or have the sensi

32、tivity validatedbefore each initial use.9.2 Calibration/Sensitivity Validation EquipmentUse thefollowing equipment for calibration of the test system:NOTE 1This equipment serves a dual function; either to calibrate theultrasonic instrument for leakage rate approximation as in 11.3,ortoverify the sen

33、sitivity for detection and location as in 11.1.9.2.1 Leak Standard, with a preset flow rate of 6.7 105mol/s (1.5 std. cm3/s at 0C) 65 % for both Class I and ClassII instruments. The orifice size shall be approximately 0.2 mm0.008 in.9.2.2 Regulator, for the nitrogen supply with output pressureand fl

34、ow gages. The tank pressure gage is optional.9.3 Air Probe Sensitivity Validation:9.3.1 For Class I instruments, locate the detection probe adistance of 10.0 m (60.1 m) from the calibrated leak standardspecified in 9.2.1. For Class II instruments, place the detectionprobe a distance of 5.0 m (60.1 m

35、) from the calibrated leakspecified in 9.2.1.9.3.2 Check to see that the detector probe and leak sourceare aligned to obtain the peak response (see Fig. 1).E1002 11 (2018)29.3.3 A signal level indication above normal backgroundfluctuations should be observed.9.3.4 Place a sound absorbing barrier in

36、front of themicrophone, blocking out the calibrated leak source; the meterreading should zero with a corresponding absence of an audiblesignal.9.4 Recalibration:9.4.1 Recheck or validate the sensitivity of the equipment atthe beginning of each shift or designated work period interval.This test must

37、be performed at the same sensing frequency asthe initial test.9.4.2 Recheck the equipment when abnormalities are ob-served in its operation.9.5 Ultrasonic Transmitter Sensitivity Validation, shouldhave the generated amplitude validated before each initial use.This could be done by placing the ultras

38、onic transmitter a fixeddistance from the detection instrument to obtain a reading or bytaking a NIST-traceable reading with a sound pressure leveldetector. Another method would be to place the ultrasonictransmitter in a container with a known leak that is equivalentto the leaks that are to be detec

39、ted. To eliminate vessel orcontainer vibrations from influencing the readings, the interiorshould be lined with closed cell foam, or rubber, or both. Theorifice size shall be equivalent to the leaks that are to bedetected.10. General Considerations10.1 Openings:10.1.1 Seal all openings using plugs,

40、covers or other suit-able materials that can be readily and completely removed afterthe completion of the test.10.1.2 Provide a gas inlet by attaching a valve to one of thetest covers on all items to be pressurized.10.2 Check of Test Parts:10.2.1 Examine the part and test equipment before pressureis

41、 applied to ensure that it is tight and all appurtenances thatshould not be subjected to the test pressure have been discon-nected or isolated by valves or other suitable means.10.2.2 Check safe pressure rating to be sure it will not beexceeded during test.10.3 Temperature of Vessel and Testing Medi

42、um:10.3.1 The temperature of the pressurizing gas must not beat a level that would be injurious to the part or its components.10.3.2 The test gas is dry nitrogen. (Compressed air cancontain oil or water droplets which can seal leaks.)10.4 Pressure:10.4.1 Unless otherwise specified the gage pressure

43、shall beat least 70 kPa 10 psig. Normal safety precautions should beobserved when pressurizing the part under test with the gas toavoid test part ruptures.10.4.2 Gradually increase the pressure in the part to finaltest pressure.10.5 Ultrasonic Transmitter MethodInspect test object tobe sure it is fr

44、ee of debris and liquids at the test sites.11. Detection and Location of Pressurized Gas Leaks11.1 Detection:11.1.1 Set the sensitivity at maximum.11.1.2 Begin to scan by pointing the probe towards the testarea. The procedure is to go from “high sensitivity” to “lowsensitivity,” reducing the sensiti

45、vity as the leak is approached.FIG. 1 Calibrated Leak Standard/Sensitivity Validation EquipmentE1002 11 (2018)311.1.3 Note the fluctuations in meter readings and thevolume from the earphones or speaker.11.1.4 If there is too much ultrasound in the area, reduce thesensitivity and continue to scan.11.

46、1.5 If it is difficult to isolate the leak due to competingultrasound, place the focusing probe accessory over the scan-ning probe. This increases the directional response character-istics of the probe.11.1.6 Listen for a “rushing” sound while observing themeter.11.1.7 Follow the ultrasound to the l

47、oudest point. The meterwill show a higher reading as the leak is approached.11.2 Ultrasonic Transmitter MethodProvide for access oftransmitter placement in the test object or provide an inlet pipeconduit fitting to couple with the transmitter allowing it toadequately saturate the test object with ul

48、trasound in such away as to shield stray ultrasonic leakage into the ultrasonic leakdetection side.11.2.1 All potential interferences shall be cleared away fromthe test areas either by cleaning or by blowing the areas drywith compressed gas.11.2.2 Transmitter PlacementThe transmitters shall beplaced

49、 to adequately cover the test object with a uniformultrasound.11.2.3 Listen for the distinct heterodyned ultrasound pro-duced by the ultrasonic transmitter. This may be a whistle orwarbling tone.11.3 Location:11.3.1 In order to focus on the leak, keep reducing thesensitivity until a leak is located.11.3.2 To Confirm a LeakPosition the scanning probe withor without the rubber focusing extension close to the suspectleak site and move it, slightly back and forth, up and down. Ifthe leak is at this location, the sound and meter readings

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1