ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf

上传人:eastlab115 文档编号:533797 上传时间:2018-12-06 格式:PDF 页数:6 大小:74.86KB
下载 相关 举报
ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf_第1页
第1页 / 共6页
ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf_第2页
第2页 / 共6页
ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf_第3页
第3页 / 共6页
ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf_第4页
第4页 / 共6页
ASTM E932-1989(2002) Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers《扩散式红外线分光计性能描述和测量的标准实施规范》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: E 932 89 (Reapproved 2002)Standard Practice forDescribing and Measuring Performance of DispersiveInfrared Spectrometers1This standard is issued under the fixed designation E 932; the number immediately following the designation indicates the year oforiginal adoption or, in the case of r

2、evision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the necessary information toqualify dispersive infrared instruments for spe

3、cific analyticalapplications, and especially for methods developed by ASTMInternational.1.2 This practice is not to be used as a rigorous test ofperformance of instrumentation.1.3 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsib

4、ility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 131 Terminology Relating to Molecular Spectroscopy2E 168 Practices for General Techniques of In

5、frared Quanti-tative Analysis2E 387 Test Method for Estimating Stray Radiant PowerRatio of Spectrophotometers by the Opaque Filter Method2E 1252 Practice for General Techniques for Obtaining In-frared Spectra for Qualitative Analysis23. Terminology3.1 Definitions and SymbolsFor definitions of terms

6、andsymbols, refer to Terminology E 131 and Compilation ofASTM Standard Definitions.34. Significance and Use4.1 This practice is intended for all infrared spectroscopistswho are using dispersive instruments for qualitative or quan-titative areas of analysis.4.2 The purpose of this practice is to set

7、forth performanceguidelines for testing instruments used in developing ananalytical method. These guidelines can be used to compare aninstrument in a specific application with the instrument(s) usedin developing the method.4.3 An infrared procedure must include a description of theinstrumentation an

8、d of the performnace needed to duplicate theprecision and accuracy of the method.5. Apparatus5.1 For the purposes of this practice, dispersive instrumentsinclude those employing prisms, gratings, or filters to separateinfrared radiation into its component wavelengths.5.2 For each new method, describ

9、e the apparatus andinstrumentation both physically and mechanically, and also interms of performance as taught in this practice. That is, thedescription should give numerical values showing the fre-quency accuracy and the frequency and the photometricprecision. State the spectral slit width maximum

10、or slit widthprogram if one is used. Where possible, state the maximum andminimum resolution if those data are a part of the instrumentdisplay. Show typical component spectra as produced by theinstrument to establish the needed resolution.5.3 If a computer program is used, describe the program.Inclu

11、de the programming language and availability, or whetherthe program is proprietary to a manufacturer.6. Reference to this Practice in Standards6.1 Reference to this practice should be included in allASTM infrared methods. The reference should appear in thesection on apparatus where the particular sp

12、ectrometer isdescribed.7. Parameters in Spectroscopy7.1 Dispersive infrared spectrometers have a source ofquasi-monochromatic radiation together with a photometer formeasuring relative radiant power. Accurate spectrometry in-volves a large number of interrelated factors that determine thequality of

13、the radiant power passing through a sample and thesensitivity and linearity with which this radiant power can bemeasured. Assuming proper instrumentation and its use, theinstrumental factors responsible for inaccuracies in spectrom-etry are resolution, linearity (Practices E 168), stray radiantpower

14、 (Test Method E 387), and cell constants (PracticeE 1252). Rigorous measurement of these factors is beyond thescope of this practice, and a more practical approach isdescribed for the accessible factors.1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and is th

15、e direct responsibility of Subcommittee E13.03 on InfraredSpectroscopy.Current edition approved Aug. 25, 1989. Published October 1989.2Annual Book of ASTM Standards, Vol 03.06.3Available from ASTM International Headquarters, 100 Barr Harbor Drive,West Conshohocken, PA 19428.1Copyright ASTM Internati

16、onal, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.8. Instrument Operation8.1 The analyst selects the proper instrumental operatingconditions in order to get satisfactory performance (1-3).4Because instrument design varies, the manufacturers recom-mendations ar

17、e usually best. A record of operating conditionsshould be kept so that data can be duplicated by future users.8.2 In addition to operating conditions, the following shouldbe checked and recorded:8.2.1 Ambient temperature,8.2.2 Pen response time,8.2.3 Scanning speed,NOTE 1In some instruments these fu

18、nctions are integrated in the scanmodes.8.2.4 Noise level, and8.2.5 Mechanical repeatability.8.3 Each of the above factors is important in the measure-ment of analytical wavenumber and photometric data. There isusually some lag between the recorded reading and the correctreading. Proper selection of

19、 operating conditions and good,reproducible, sample handling techniques minimize these ef-fects or make the effects repeatable. For example:8.3.1 Variation in temperature of the monochromator orsample may cause changes in wavenumber precision andaccuracy.8.3.2 Scanning too fast will displace the app

20、arent wavenum-ber towards the direction scanned and will decrease the peakabsorbance reading for each band.NOTE 2Some instruments provide for automatic monitoring andcorrection of this effect.8.4 Mechanical repeatability of the monochromator andrecording system as well as positioning of chart paper

21、areimportant in wavenumber measurement.8.4.1 Chart paper should be checked for uniformity of theprinted scale length as received and rechecked at time of use,particularly if the paper has been subjected to pronouncedhumidity changes. Instructions on obtaining proper mechanicalrepeatability may be gi

22、ven in the manufacturers literature.8.5 In the case of computerized dispersive instruments, anyspectrum printed from a computer file must be obtained asprescribed by the manufacturer and should be identical to theoriginal data.PRECISION AND ACCURACY9. Definitions9.1 wavenumber precisiona measure of

23、the capability of aspectrometer to return to the same spectral position as mea-sured by a well-defined absorption or emission band when theinstrument is reset or rescanned. The index used in this practiceis the standard deviation.9.2 wavenumber accuracythe deviation of the averagewavenumber reading

24、of an absorption band or emission bandfrom the known wavenumber of that band.10. Nature of Test10.1 For the purpose of calibration, most methods employpure compounds and known mixtures at specified analyticalwavenumbers. The wavenumbers are either read from a dial,optical display, chart paper, or a

25、computer file.11. Reference Wavenumbers in the Infrared Region (2)11.1 The recommended wavenumber calibration points arethe absorption maxima of a standard (98.4/0.8/0.8 by weight)indene/camphor/cyclohexanone mixture listed in Table 1. Suit-able path lengths are 0.2 mm for the range from 3800 to 158

26、0cm1and 0.03 mm for the wavenumber range from 1600 to 600cm-1. A mixture containing equal parts by weight of indene,camphor, and cyclohexanone (1/1/1 by weight) at a path lengthof 0.1 mm may be used for the range from 600 to 300 cm1.See Table 2 and Fig. 1.11.2 Polystyrene is also a convenient calibr

27、ation standardfor the wavenumber range from 4000 to 400 cm1. Polystyrenefilms, approximately 0.03 to 0.05 mm thick, can be purchasedfrom instrument manufacturers. The recommended calibrationpeaks are listed in Fig. 2.NOTE 3The correction of frequency for the refractive index of air issignificant in

28、the wavenumber calculation only when wavelengths havebeen measured to better than 3 parts in 10 000. Reference (3) tabulatesadditional reference wavenumbers of interest.11.3 For low-resolution prism or filter instruments operatedin single-beam mode, the position of the atmospheric carbondioxide band

29、 near 2350 cm1can be useful. This band may beresolved into a doublet. The 2350-cm1value is for theapproximate center between the two branches. The atmo-spheric carbon dioxide band near 667 cm1is useful in thelow-wavenumber region.12. Dynamic Error Test12.1 For dispersively measured spectra, the foll

30、owing dy-namic error test is suitable for use with most grating and prism4The boldface numbers refer to the list of references at the end of this practice.TABLE 1 Indene-Camphor-Cyclohexanone (98.4/0.8/0.8)MixtureRecommended Calibration Bands5BandNo.Wavenumber,cm1BandNo.Wavenumber,cm11 3927.2 6 1.0

31、44a 1741.92 3901.6 44b 1713.43 3798.9 47 1661.85 3660.6 6 1.0 48 1609.88 3297.8 6 1.0 49 1587.59 3139.5 51 1553.210 3110.2 53 1457.3 6 1.012 3025.4 54 1393.515 2887.6 55 1361.117 2770.9 57 1312.419 2673.3 58 1288.020 2622.3 60 1226.221 2598.4 6 1.0 61 1205.123 2525.5 62 1166.128 2305.1 64 1122.429 2

32、271.4 66 1067.7 6 1.030 2258.7 67 1018.533 2172.8 69 947.234 2135.8 6 1.0 70 942.435 2113.2 71 914.736 2090.2 72 861.339 1943.1 73 830.540 1915.3 74 765.341 1885.1 76 718.142 1856.9 77 692.6 6 1.044 1797.7 6 1.0E 9322spectrometers. (4 and 5)12.2 The spectrum of the (98.4/0.8/0.8) indene/camphor/cycl

33、ohexanone mixture is remeasured from 1350 to 850 cm1at one fourth of the scan rate used for the reference spectrumand with other operating conditions unchanged. The heightsfrom the baseline of the bands at 1288.0, 1226.2, 1205.1,1018.5 and 914.7 cm1are measured in absorbance units onboth the fast an

34、d slowly scanned charts. The absorbance ratiosA1226.2/A1288.0, A1205.1/A1226.2, and A914.7/A1018.5should not differ by more than6 0.02 between the fast andslow runs.NOTE 4The indene/camphor/cyclohexanone should remain in sealed,refrigerated ampoules.13. Selection of Slit Width or Slit Program13.1 On

35、e of the most important parameters the analyst mustselect is the spectral slit width. The slit width affects resolutionand the signal-to-noise ratio (S/N). Generally, a narrower slitwidth gives higher resolution and lower S/N ratio. These mustbe optimized for any given analysis.13.2 The preferred ma

36、nner of expressing resolution is interms of spectral band width, but methods of measuring thisquantity in all spectral regions are not available.13.2.1 Spectral band width is not constant throughout thespectrum and therefore must be determined in each region ofinterest. In the neighborhood of 1200 c

37、m1, the spectral bandwidth can be determined approximately from the ratioA1205.1/A1226.2 of the (98.4/0.8/0.8) indene/camphor/cyclohexanone mixture, computed in the dynamic error test, asgiven in Table 3.13.3 In each infrared method, typical spectra of the compo-nents, or a spectrum of a suitable mi

38、xture of components,should be included to illustrate the resolution found to beadequate to perform the analysis. These spectra should bedirect copies of the plotted spectra and not redrawn curves.PHOTOMETRY14. Linearity of Absorbance14.1 In a spectrometric method, photometric data are usedto determi

39、ne concentrations. Linearity of absorbance is afunction of instrument response. The relationship must bedetermined in the concentration range of interest.14.2 Procedure for testing linearity and establishing work-ing curves are described in Practices E 168.14.3 Some methods for quantitative analysis

40、 do not requirelinear response. The ultimate criterion for these is whether amethod gives correct answers for known samples.15. Measurement Procedure for Frequency Accuracy andPrecision15.1 From Tables 1-3, select calibration wavenumbers,preferably bracketing each analytical wavenumber and readeach

41、wavenumber ten times.15.2 Average the observed readings for each wavenumber.The wavenumber accuracy is the difference between the truewavenumber and the average observed wavenumber.NOTE 5To check the wavenumber accuracy of a non-scanning instru-ment, balance the instrument at the true value of the a

42、bsorbancemaximum. Adjust the wavenumber drive until maximum apparent absor-bance is found. Always approach the line or band from the same direction.Repeat ten times.15.3 Calculate the precision of each observed wavenumberusing the following equation:s 5(Xi2 X!2n 2 1(1)where:s = estimated standard de

43、viation of the series of results,Xi= individual observed value (wavenumber, absorbance,or transmittance)X= average (arithmetic mean) of the observed values, andn = number of observations.15.4 Results should be specified in the following order: truepeak position of reference material, average wavenum

44、berdetermined, and wavenumber standard deviation.16. Measurement Procedure for Photometric Precision16.1 Photometric precision represents the capability of thephotometer system to reproduce the same value in successivedeterminations. The index of precision used in this practice isthe standard deviat

45、ion.16.2 Photometric precision is determined on a calibrationsample by measuring the absorbance or transmittance of thesame sample ten times, following the same procedure used toobtain the data for the linearity test.16.3 Tabulate the individual readings of apparent absor-bance or transmittance. Cal

46、culate the standard deviation of theten readings using the equation in 15.3. Report the averagereading and the standard deviation.17. Stray Radiate Power17.1 Stray radiant power (SRP) causes an error in thespectrometer zero transmittance reading. If measured at thedesignated path lengths, the indene

47、/camphor/cyclohexanonereference spectrum should show essentially zero transmittanceat 3050.0, 1609.6 and 765.4 cm1. The test spectra at thesewavenumbers should therefore match the spectrometer trans-mittance zero within the manufacturers tolerances. A0.4-mmlayer of pure indene is almost totally abso

48、rbing at 392, 420, and551 cm1and this can be used to establish the stray radiantpower below 600 cm1. This should not exceed 2 % transmit-tance at wavenumbers greater than 600 cm1; the permissibleamount at lower wavenumbers is left to the discretion of theevaluator.TABLE 2 Indene-Camphor- Cyclohexano

49、ne (1/1/1) MixtureRecommended Calibration Bands5Band No.Wavenumber,cm11 592.12 551.73 521.44 490.2 6 1.05 420.56 393.17 381.68 301.4E 932318. Performance Evaluation18.1 Performance is adequate when test results are equiva-lent to tests on the instrument(s) used in developing the specificmethod.519. Keywords19.1 infrared spectroscopy; molecular spectroscopy5Table 1 and Table 2 contain wavenumber values for bands in Fig. 1.NOTE 1See Table 1 and Table 2 for cm1of numbered absorption maxima.FIG. 1 IUPAC Definitive Spectra of Indene-Camphor-Cyclohexanone Mixtu

展开阅读全文
相关资源
猜你喜欢
  • BS ISO 6846-1995 en_2111  Photography - Black and white continuous tone papers - Determination of ISO speed and ISO range for printing《摄影技术.黑白连续色调相纸.晒印用ISO感光度和ISO幅距的测定》.pdf BS ISO 6846-1995 en_2111 Photography - Black and white continuous tone papers - Determination of ISO speed and ISO range for printing《摄影技术.黑白连续色调相纸.晒印用ISO感光度和ISO幅距的测定》.pdf
  • BS ISO 6849-1997 en_4460  Photography - Processing wastes - Determination of boron《摄影技术.冲洗废液.硼的测定》.pdf BS ISO 6849-1997 en_4460 Photography - Processing wastes - Determination of boron《摄影技术.冲洗废液.硼的测定》.pdf
  • BS ISO 6851-2002 en_7679  Photography - Processing waste - Determination of total amino nitrogen (microdiffusion Kjeldahl method)《摄影.冲洗废液.氨态氮总含量的测定(基耶达微扩散法)》.pdf BS ISO 6851-2002 en_7679 Photography - Processing waste - Determination of total amino nitrogen (microdiffusion Kjeldahl method)《摄影.冲洗废液.氨态氮总含量的测定(基耶达微扩散法)》.pdf
  • BS ISO 6853-2002 en_5663  Photography - Processing waste - Determination of ammoniacal nitrogen (microdiffusion method)《摄影.冲洗废液.氨态氮含量的测定(微扩散法)》.pdf BS ISO 6853-2002 en_5663 Photography - Processing waste - Determination of ammoniacal nitrogen (microdiffusion method)《摄影.冲洗废液.氨态氮含量的测定(微扩散法)》.pdf
  • BS ISO 6855-1-2012 en_5608  Mopeds Measurement method for gaseous exhaust emissions and fuel consumption General test requirements《轻便摩托车.气体废气排放和燃料消耗的测量方法.通用试验要求》.pdf BS ISO 6855-1-2012 en_5608 Mopeds Measurement method for gaseous exhaust emissions and fuel consumption General test requirements《轻便摩托车.气体废气排放和燃料消耗的测量方法.通用试验要求》.pdf
  • BS ISO 6855-2-2012 en_2618  Mopeds Measurement method for gaseous exhaust emissions and fuel consumption Test cycles and specific test conditions《轻便摩托车.气体排放和燃料消耗的测量方法.试验循环和特效试验条件 n n》.pdf BS ISO 6855-2-2012 en_2618 Mopeds Measurement method for gaseous exhaust emissions and fuel consumption Test cycles and specific test conditions《轻便摩托车.气体排放和燃料消耗的测量方法.试验循环和特效试验条件 n n》.pdf
  • BS ISO 6855-3-2012 en_6661  Mopeds Measurement method for gaseous exhaust emissions and fuel consumption Fuel consumption measurement at a constant speed《助力车 废气排放和燃料消耗的测量方法 在恒速条件下燃料消耗的测量》.pdf BS ISO 6855-3-2012 en_6661 Mopeds Measurement method for gaseous exhaust emissions and fuel consumption Fuel consumption measurement at a constant speed《助力车 废气排放和燃料消耗的测量方法 在恒速条件下燃料消耗的测量》.pdf
  • BS ISO 6856-2005 en_2859  Road vehicles - Unscreened high-voltage ignition cable assemblies - Test methods and general requirements《道路车辆.非屏蔽高压点火电缆组件.试验方法和一般要求》.pdf BS ISO 6856-2005 en_2859 Road vehicles - Unscreened high-voltage ignition cable assemblies - Test methods and general requirements《道路车辆.非屏蔽高压点火电缆组件.试验方法和一般要求》.pdf
  • BS ISO 6861-1997 en_5416  Information and documentation - Glagolitic alphabet coded character set for bibliographic information interchange《信息和文件管理.文献信息交换用格拉哥里字母代码字符集》.pdf BS ISO 6861-1997 en_5416 Information and documentation - Glagolitic alphabet coded character set for bibliographic information interchange《信息和文件管理.文献信息交换用格拉哥里字母代码字符集》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1