资源描述

1、A-A-5473b m 9999974 OLL9bBO 749 m A-A-54 7 3 6 23 November 1992 COMMERCIAL ITEM DESCRIPTION THERMOMETER, CLINICAL, HUMAN (Electronic, Battery Powered, With Battery Charger) The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Descriptio

2、n covers requirements for an electronic thermometer suitable for oral or rectal body temperature measurement in humans. Salient characteristics: The thermometer shall be a portable, battery powered, Fahrenheit/Celsius electronic clinical thermometer system consisting of a thermometer, battery charge

3、r/storage base, oral probe, rectal probe, disposable probe covers, and dispenser probe cover brackets. The battery charger shall operate from 110/220V, 50/60 Hz power. The battery charger shall be in compliance with UL 544, Standard for Medical and Dental Equipment. The thermometer shall have a pred

4、ictive mode temperature range of 90.0F to 107.gF, plus or minus 0.2OF. The thermometer shall have monitor mode temperature range of 88.0F to 107.gF, plus or minus O.ZF. The thermometer shall have a pulse mode range of 20 to 200 beats per minute (BPM). Pulse mode activation shall be by depressing a b

5、utton on the bottom of the thermometer. The thermometer shall have an LED charge light indicator on the display panel when the battery is recharging. AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or ne

6、tworking permitted without license from IHS-,-,-A-A-5 4 7 3 6 The thermometer shall have an audible tone that sounds when temperature has been computed, when pulse has been calculated, and when an error message has occurred. The thermometer shall have a display panel that indicates temperature readi

7、ngs, tissue contact, pulse rate, and error messages on a five character minimum LED display. A rechargeable nickel-cadmium battery, that automatically recharges when placed in the battery charger/storage base shall be supplied with the thermometer. The battery shall recharge in no more than 10 hours

8、 Battery life shall be at least five hours continuous from fully charged state. The thermometer shall have the capability of both Celsius and Fahrenheit readings. The thermometer shall be capable of being hung around the neck for convenience during use. The thermometer shall have a compartment for

9、holding a box of disposable probe covers. The thermometer shall include one oral probe with a spring action cord approximately 2 feet long when coiled, approximately 4 feet long when stretched. It shall be capable of attachment to and detachment from the thermometer with ease. One rectal probe with

10、a spring action cord approximately 2 feet long when coiled, approximately 4 feet long when stretched shall also be included with the thermometer. It shall be capable of attachment to and detachment from the thermometer with ease. The thermometer shall be provided with 1000 clean, disposable, rigid p

11、lastic probe covers. The covers shall be 6 inches long and shall be fitted onto and removed from each probe without having to touch the probe cover. The thermometer shall be provided with a probe cover dispenser bracket which may be metal or plastic. If metal, it shall be corrosion-resisting steel,

12、properly hardened and passivated. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A - A-5 4 7 3 b W 9999974 0119682 511 W A-A-54 7 3 6 Wor1ririanr;liip. The electronic thermometer shall be free from defects which detract from its appearance or impa

13、ir its serviceability. Unit. Each (EA). Oiie electronic thermometer, as specified, constitutes one unit. Quality Assurance Provisions. Responsibility for inspection. Unless otherwise specifiec in the contract or purchase order, the contractor is responsible for the performance of all inspection requ

14、irements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to per

15、form any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Records. Records of examinations and tests perfcrmed by or for the contractor shall be maintained by the contractor and made avai

16、lable to the Government upon the Governments request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. Inspection. Inspection, as used herein, is defined as both examination (such as

17、 visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item. Samplincr for examination. Sampling for examination shall be conducted in accordance with MIL-STD-105, wit

18、h an AQL of 1.0 (percent defective) and an inspection level of II. The unit of product for examination purposes shall be one thermometer. Samplins for tests. Sampling for tests, including dimensional test, shall be conducted in accordance with MIL-STD-105, with an AQL of 1.0 (percent defective) and

19、an inspection level of S-2. shall be one thermometer. The unit of product for test purpos S 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54736 H 9iY OLL9683 458 A-A-54 7 3 6 Tests. Tests shall be conducted to determine compliance with specif

20、ication requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics. Tests shall include, but shall not be limited to, the following: Accuracy test. Set the thermometer in the Monitor mode and set the Fahrenheit (OF) = Celsius (OC) to OF or O

21、C as desired. Install probe cover on the probe. Insert probe tip into a circulating water bath that has a known temperature between 88OF and 107.9OF and is accurate to within plus or minus 0.03OF. After 3 to 5 minutes, the displayed temperature should completely stop changing and shall be e ual to k

22、nown bath temperature to within plus or minus 0.2 8 F. Performance test. The thermometer shall be subjected to appropriate tests to determine that the thermometer conforms to all performance requirements of this specification. Metric products. Products manufactured to metric dimensions will be consi

23、dered on an equal basis with those manufactured using inch-pound units, providing they fall within the tolerances specified using conversion tables contained in the latest revision of Federal Standard 376, and all other requirements of this document are met. If a product is manufactured to metric di

24、mensions and those dimensions exceed the tolerances specified in the inch/pound units, a request should be made to the contracting officer to determine if the product is acceptable. The contracting officer has the option of accepting or rejecting the product. Contractor certification. The contractor

25、 shall certify and maintain substantiating evidence that the product offered meets the salient characteristics of this Commercial Item Description, and that the product conforms to the producers own drawings, specifications, standards, and quality assurance practices. The government reserves the rig

26、ht to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract. 4 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5473b 9999974 OLL9b84 394 = A-A-54 7 3 6 Requl

27、atory requirements. Federal Food, Drus and Cosmetic Act. If the product covered by this document has been determined by the U.S. Food and Drug Administration to be under its jurisdiction, the of feror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with th

28、e requirements of the Federal Fuod, Drug and Cosmetic Act, as amended, and regulations promulgated thereunder. In addition, the offeror/contractor shall comply, and be responsible for conipliarice by its subcontractors/suppliers, with the requirements of all oLher applicable Federal, State, and loca

29、l statutes, ordinances, and regulations. Recovered materials. The offeror/contractor is encouraged to use recovered material in accordance with Federal Acquisition Regulation Subpart 23.4 to the maximum extent practical. Preservation, packaqinq, D ackinq, labelinq, and markinq. Unless otherwise spec

30、ified, preservation, packaging, and packing shall be to a degree of protection to preclude damage to containers and/or contents thereof under normal shipping conditions, handling, etc:, involving shipment from the supply source to the receiving activity, plus reshipment from receiving activity, and

31、shall conform to applicable carriers rules and regulations. Intermediate and exterior package quantities and labeling and marking shall be as specified in the contract and/or order. Orderins data. Intermediate/exterior package quantities, labeling, and marking must be specified in the contract and/o

32、r order. NOTE: The following National Stock Number is covered by this document. 6515-01-187-6606 MILITARY INTERESTS: PREPARING ACTIVITY: Custodians: DoD-MB Air Force - 03 Army - MD Navy - MS 5 Agent: DLA-DM Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A - A-5 4 7 3 b 9999974 OLL9685 220 A-A-54 7 36 CIVIL AGENCY COORDINATING ACTIVITIES: F UA- MPQAS VA - OSS USPHS Project Number: 6515-5343 Location: THERM17C.DOC 6 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-

展开阅读全文
相关资源
猜你喜欢
  • DIN 11746-1999 Agricultural machinery - Determination of playload vertical static load axle load - Rigid drawbar trailer drawbar trailer《农业机械 有效载荷 垂直静态载荷 轴向载荷的测定 刚性牵引杆拖车和牵引杆拖车》.pdf DIN 11746-1999 Agricultural machinery - Determination of playload vertical static load axle load - Rigid drawbar trailer drawbar trailer《农业机械 有效载荷 垂直静态载荷 轴向载荷的测定 刚性牵引杆拖车和牵引杆拖车》.pdf
  • DIN 11747-1992 Agricultural tractors and machinery leaf springs for trailers dimensions《农用拖拉机和机具 挂车板簧 尺寸》.pdf DIN 11747-1992 Agricultural tractors and machinery leaf springs for trailers dimensions《农用拖拉机和机具 挂车板簧 尺寸》.pdf
  • DIN 11751-1989 Agricultural tractors and machinery hydraulic cylinders for trailers connecting dimensions《农用拖拉机和机械 拖挂车用液压筒(200巴) 连接尺寸》.pdf DIN 11751-1989 Agricultural tractors and machinery hydraulic cylinders for trailers connecting dimensions《农用拖拉机和机械 拖挂车用液压筒(200巴) 连接尺寸》.pdf
  • DIN 118-1-1977 Driving Elements Pedestal Plain Bearings for General Mechanical Engineering Applications Main Dimensions《传动元件 一般机械工程用托架滑动轴承 主要尺寸》.pdf DIN 118-1-1977 Driving Elements Pedestal Plain Bearings for General Mechanical Engineering Applications Main Dimensions《传动元件 一般机械工程用托架滑动轴承 主要尺寸》.pdf
  • DIN 1180-1971 Drain pipes of clay dimensions general quality requirements and test methods《粘土排水管 尺寸、一般质量要求和试验方法》.pdf DIN 1180-1971 Drain pipes of clay dimensions general quality requirements and test methods《粘土排水管 尺寸、一般质量要求和试验方法》.pdf
  • DIN 11800-1999 Agricultural engineering - Electronic data interchange in agriculture - Data interchange in the field of dairy farming《农业机械 农业中的电子数据交换 奶畜饲养业中的数据交换》.pdf DIN 11800-1999 Agricultural engineering - Electronic data interchange in agriculture - Data interchange in the field of dairy farming《农业机械 农业中的电子数据交换 奶畜饲养业中的数据交换》.pdf
  • DIN 11832-1-1990 Farm and livestock husbandry equipment shut-off valves for slurry pipes gate valves for static pressure of max 1 bar《农场和牲畜饲养设备 输送液态粪肥的管道用阀门管件 第1部分 最大1巴的静压力闸阀》.pdf DIN 11832-1-1990 Farm and livestock husbandry equipment shut-off valves for slurry pipes gate valves for static pressure of max 1 bar《农场和牲畜饲养设备 输送液态粪肥的管道用阀门管件 第1部分 最大1巴的静压力闸阀》.pdf
  • DIN 1184-1-1992 Pumping stations design construction and operation《泵站 设计 建造和运行》.pdf DIN 1184-1-1992 Pumping stations design construction and operation《泵站 设计 建造和运行》.pdf
  • DIN 1184-2-1992 Pumping stations centrifugal pumps for operation with screen directives for planning《泵站 带滤网运行的离心泵 设计用指导性文件》.pdf DIN 1184-2-1992 Pumping stations centrifugal pumps for operation with screen directives for planning《泵站 带滤网运行的离心泵 设计用指导性文件》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1