1、Designation: F 700 93 (Reapproved 2008)Standard Practice forCare and Handling of Intracranial Aneurysm Clips andInstruments1This standard is issued under the fixed designation F 700; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision,

2、the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers recommended procedures for han-dling of neurosurgical intracranial aneurysm clips an

3、d the clipappliers (instruments).1.2 This practice is intended to inform hospital receivingpersonnel, central supply personnel, operating room personnel,and other individuals who will handle intracranial aneurysmclips and the instruments related thereto of recommended careand handling procedures to

4、prevent damage of intracranialaneurysm clips and instruments.1.3 Handling and packaging procedures for the product arenot a part of this practice and are covered in other practices.2. Terminology2.1 Definitions:2.1.1 clip applierany clip holder designed specifically fora particular type clip used du

5、ring surgical procedures involvingthe implantation of intracranial aneurysm clips. This device isreferred to in this practice as a clip applier.2.1.2 intracranial aneurysm clipa device introduced sur-gically to occlude the blood inlet into an intracranial aneurysmwith the intention that it remain wi

6、thin the body following thesurgery. This device is referred to in this practice as an“implant,” specifically as an intracranial aneurysm clip.3. Receiving Implants and Instruments3.1 Receipt:3.1.1 Many implants are wrapped in special packages,envelopes, or other containers. These wrappings should no

7、t beremoved by the receiving personnel.3.1.2 Carefully handle nonsterilized implants and instru-ments upon receipt to avoid scratching, marking, or abrasionby other implants, instruments, unpacking tools, or by drop-ping or otherwise endangering the surface finish or configura-tion.3.2 TransportTran

8、sport in a manner to preclude anydamage or alteration to the received condition of the implant orinstrument (clip applier).3.3 Storage:3.3.1 Store the aneurysm clips and appliers prior to use insuch a manner to maintain the devices surface finish orconfiguration, or both.3.3.2 Implants will be ident

9、ified by a unique code on thesurface of the device. When implanted, it is necessary that thetype of aneurysm clip and its code be noted in the operativereport.3.3.3 Stock RotationThe principle of first in, first out, isrecommended.3.3.4 Store implants in the operating room in such a manneras to isol

10、ate and protect the implants surface, sterility, andconfiguration. Keep implants made of different metals sepa-rated.4. Cleaning and Sterilization4.1 Prior to resterilization and promptly following eachsurgical procedure, thoroughly and carefully clean all instru-ments and implants with approved cle

11、aning techniques. Ultra-sonic cleaners or hand scrubbing are suitable methods ifcarefully done. The method employed should be utilized toprevent impact, scratching, bending, or surface contact withany materials that might affect the implant or instrumentconfiguration. Aneurysm clips must not be open

12、ed except bytheir specific applier.4.2 Any clip that has been implanted or in direct contactwith blood or body fluids and is not used in the procedureshould not be reused in any subsequent procedure. The clipsshould be discarded; in accordance with tracing requirements,the manufacturer must be infor

13、med of each clip that isdiscarded.4.3 After cleaning, rinse the neurosurgical intracranial an-eurysm clips and instruments completely free of all residuals1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of Su

14、bcommitteeF04.31 on Neurosurgical Standards.Current edition approved Feb. 1, 2008. Published March 2008. Originallyapproved in 1981. Last previous edition approved in 2002 as F 700 93 (2002).1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United S

15、tates.including soap, detergent, or cleaning solutions; and drythoroughly. Devote special attention to hinges on the instru-ments and recesses on the clips since these are points thatentrap both chemicals and rinse water. In so doing, the clipmust not be opened except by its designated applier.4.4 L

16、ubricate clip appliers that require lubrication immedi-ately after drying. Follow the recommendations of the manu-facturers of such instruments explicitly as to the method, type,and amount of lubricant. Do not lubricate the clips.4.5 Carry out sterilization by steam autoclaving or othermethods in a

17、manner that protects the integrity of the aneurysmclips and appliers.4.6 Do not sterilize implants in contact with instruments orimplants of other materials. Metallic oxide could transfer to theimplant, initiating an unacceptable condition.5. Use of Neurosurgical Intracranial Aneurysm Clips andAppli

18、ers5.1 Open and close an aneurysm clip with the clip applierspecifically manufactured to handle that particular aneurysmclip. At no time should an aneurysm clip be opened with thefingertips.5.1.1 Any aneurysm clip that has been opened with thefingertips should be discarded and never used in a patien

19、t.5.2 No attempt should be made to alter the configuration ofpreformed aneurysm clip legs or springs.5.2.1 Dispose of intracranial aneurysm clips that exhibitsurface or configuration damage.5.2.2 Dispose of intracranial aneurysm clips whose perfor-mance capabilities have been jeopardized by mishandl

20、ing orimproper care.6. Keywords6.1 handling of implants; instruments; material handling;neurosurgical medical/devices applicationsASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard a

21、re expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not

22、 revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which y

23、ou may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org).F 700 93 (2008)2

展开阅读全文
相关资源
猜你喜欢
  • BS PD IEC TS 62763-2013_5284 Pilot function through a control pilot circuit using PWM (pulse width modulation) and a control pilot wire《通过控制导向线使用PWM (脉冲宽度调制) 的导向功能和控制导向线》.pdf BS PD IEC TS 62763-2013_5284 Pilot function through a control pilot circuit using PWM (pulse width modulation) and a control pilot wire《通过控制导向线使用PWM (脉冲宽度调制) 的导向功能和控制导向线》.pdf
  • BS ISO 8070-2007 Milk and milk products - Determination of calcium sodium potassium and magnesium contents - Atomic absorption spectrometric method《牛奶和奶制品 钙、钠、钾和镁含量的测定 原子吸.pdf BS ISO 8070-2007 Milk and milk products - Determination of calcium sodium potassium and magnesium contents - Atomic absorption spectrometric method《牛奶和奶制品 钙、钠、钾和镁含量的测定 原子吸.pdf
  • BS ISO 8082-1-2009 Self-propelled machinery for forestry - Laboratory tests and performance requirements for roll-over protective structures - General machines《林业用自推进机械 防倾.pdf BS ISO 8082-1-2009 Self-propelled machinery for forestry - Laboratory tests and performance requirements for roll-over protective structures - General machines《林业用自推进机械 防倾.pdf
  • BS ISO 8082-2-2011 Self-propelled machinery for forestry Laboratory tests and performance requirements for roll-over protective structures Machines having a rotating platf.pdf BS ISO 8082-2-2011 Self-propelled machinery for forestry Laboratory tests and performance requirements for roll-over protective structures Machines having a rotating platf.pdf
  • BS ISO 8083-2006 Machinery for forestry - Falling-object protective structures (FOPS) - Laboratory tests and performance requirements《林业机械 落体防护装置(FOPS) 实验室试验和性能要求》.pdf BS ISO 8083-2006 Machinery for forestry - Falling-object protective structures (FOPS) - Laboratory tests and performance requirements《林业机械 落体防护装置(FOPS) 实验室试验和性能要求》.pdf
  • BS ISO 8086-2004 Dairy plant - Hygiene conditions - General guidance on inspection and sampling procedures《乳品厂 卫生条件 检验和取样程序通用指南》.pdf BS ISO 8086-2004 Dairy plant - Hygiene conditions - General guidance on inspection and sampling procedures《乳品厂 卫生条件 检验和取样程序通用指南》.pdf
  • BS ISO 8096-2005 Rubber- or plastics-coated fabrics for water resistant clothing - Specification《雨衣用橡胶或塑料涂覆织物 规范》.pdf BS ISO 8096-2005 Rubber- or plastics-coated fabrics for water resistant clothing - Specification《雨衣用橡胶或塑料涂覆织物 规范》.pdf
  • BS ISO 8097-2001 Aircraft Minimum airworthiness requirements and test conditions for certified air cargo unit load devices《航空器 经认证的航空货运集装单元装置最低适航性要求和试验条件》.pdf BS ISO 8097-2001 Aircraft Minimum airworthiness requirements and test conditions for certified air cargo unit load devices《航空器 经认证的航空货运集装单元装置最低适航性要求和试验条件》.pdf
  • BS ISO 8114-1993 Textile machinery and accessories - Spindles for ring-spinning and doubling machines - List of equivalent terms《纺织机械和附件 环锭纺纱机和并线机用锭子 同义术语表》.pdf BS ISO 8114-1993 Textile machinery and accessories - Spindles for ring-spinning and doubling machines - List of equivalent terms《纺织机械和附件 环锭纺纱机和并线机用锭子 同义术语表》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1