ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf

上传人:fatcommittee260 文档编号:464264 上传时间:2018-11-27 格式:PDF 页数:9 大小:303.67KB
下载 相关 举报
ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf_第1页
第1页 / 共9页
ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf_第2页
第2页 / 共9页
ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf_第3页
第3页 / 共9页
ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf_第4页
第4页 / 共9页
ASTM C1331-2018 Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics with Broadband Pulse-Echo Cross-Correlation Method《用宽带脉冲回波互相关法测量高级陶瓷中超声波速度的标准实施规程》.pdf_第5页
第5页 / 共9页
亲,该文档总共9页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: C1331 01 (Reapproved 2012)C1331 18Standard Test Method Practice forMeasuring Ultrasonic Velocity in Advanced Ceramics withBroadband Pulse-Echo Cross-Correlation Method1This standard is issued under the fixed designation C1331; the number immediately following the designation indicates t

2、he year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This test method practice describes a proc

3、edure for measurement of ultrasonic velocity in structural engineering solids suchas monolithic ceramics, toughened ceramics, and ceramic matrix composites.1.2 This test method practice is based on the broadband pulse-echo contact ultrasonic method. The procedure involves acomputer-implemented, freq

4、uency-domain method for precise measurement of time delays between pairs of echoes returned bythe back surface of a test sample or part.1.3 This test method practice describes a procedure for using a digital cross-correlation algorithm for velocity measurement.Thecross-correlation function yields a

5、time delay between any two echo waveforms (1).21.4 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the Wo

6、rld Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:3B311 Test Method for Density of Powder Metallurgy (PM) Materials Containing Less Than Two Percent PorosityC373 Test Methods for Determination of Water Absorption and Associated Properties by

7、 Vacuum Method for Pressed CeramicTiles and Glass Tiles and Boil Method for Extruded Ceramic Tiles and Non-tile Fired Ceramic Whiteware ProductsE494 Practice for Measuring Ultrasonic Velocity in MaterialsE543 Specification for Agencies Performing Nondestructive TestingE1316 Terminology for Nondestru

8、ctive Examinations2.2 ASNT Document:Documents:4Recommended Practice SNT-TC-1A for Nondestructive Testing Personnel Qualification and CertificationANSI/ASNT CP-189 Standard for Qualification and Certification of Nondestructive Testing Personnel2.3 Military Standard:ISO Document:MIL-STD-410ISO 9712 No

9、ndestructive Testing Personnel Qualification Non-destrtuctive Testing Qualification and Certifi-cation of NDT Personnel52.4 Aerospace Industries Association Document:NAS 410 Certification and Qualification of Nondestructive Testing Personnel62.5 Additional references are cited in the text and at end

10、 of this document.1 This test method practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.06 onUltrasonic Method.Current edition approved Aug. 1, 2012Feb. 1, 2018. Published November 2012February 2018. Originally appr

11、oved in 1996. Last previous edition approved in 20072012as C1331 01 (2007).(2012). DOI: 10.1520/C1331-01R12.10.1520/C1331-18.2 The boldface numbers in parentheses refer to the list of references at the end of this test method.practice.3 For referencedASTM standards, visit theASTM website, www.astm.o

12、rg, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.4 Available from American Society for Nondestructive Testing (ASNT), P.O. Box 28518, 1711 Arlingate Ln., Columbus, OH 43228-05

13、18, http:/www.asnt.org.5 Available from Standardization Documents Order Desk, DODSSP, Bldg. 4, Section D, 700 Robbins Ave., Philadelphia, PA 19111-5098, http:/www.dodssp.daps.mil.International Organization for Standardization (ISO), ISO Central Secretariat, BIBC II, Chemin de Blandonnet 8, CP 401, 1

14、214 Vernier, Geneva,Switzerland, http:/www.iso.org.6 Available from Aerospace Industries Association of America, Inc., 1250 Eye St. NW, Washington, DC, 2005.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made

15、 to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Su

16、mmary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 back surfacethe surface of a test sample which is oppos

17、ite to the front surface and from which back surface echoes arereturned at normal incidence directly to the transducer.3.1.2 bandwidththe frequency range of an ultrasonic probe, defined by convention as the difference between the lower andupper frequencies at which the signal amplitude is 6 dB down

18、from the frequency at which maximum signal amplitude occurs.3.1.3 broadband transduceran ultrasonic transducer capable of sending and receiving undistorted signals over a broadbandwidth, consisting of a thin damped piezocrystal in a buffered probe (search unit).3.1.4 buffered probean ultrasonic sear

19、ch unit as defined in Terminology E1316 but containing a delay line, or buffer rod, towhich the piezocrystal is affixed within the search unit housing and which separates the piezocrystal from the test sample (Fig. 1).3.1.5 buffer rodan integral part of a buffered probe, usually a quartz or fused si

20、lica cylinder that provides a time delay betweenthe excitation pulse from the piezocrystal and echoes returning from a sample coupled to the free end of the buffer rod.3.1.6 cross-correlation functionthe cross-correlation function, implemented by a digital algorithm, yields a time delaybetween any t

21、wo (ultrasonic) echo waveforms. This time is used to determine velocity (1).3.1.7 dispersionvariation of ultrasonic velocity as a function of wavelength, that is, frequency dependence of velocity.3.1.8 front surfacethe surface of a test sample to which the buffer rod is coupled at normal incidence (

22、designated as testsurface in Terminology E1316.3.1.9 group velocityvelocity of a broadband ultrasonic pulse consisting of many different component wavelengths.3.1.10 test samplea solid coupon or material part that meets the constraints needed to make the ultrasonic velocitymeasurements described her

23、ein, that is, a test sample or part having flat, parallel, smooth, preferably ground or polished opposing(front and back) surfaces, and having no discrete flaws or anomalies unrepresentative of the inherent properties of the material.3.1.11 wavelength ()distance that sound (of a particular frequency

24、) travels during one period (during one oscillation), = v/f,where v is the velocity of sound in the material and where velocity is measured in cm/s, frequency in MHz, and wavelength incm, herein.3.2 Other terms or nomenclature used in this test method practice are defined in Terminology E1316.4. Sig

25、nificance and Use4.1 The velocity measurements described in this test method practice may be used to characterize material variations that affectmechanical or physical properties. This procedure is useful for measuring variations in microstructural features such as grainstructure, pore fractions, an

26、d density variations in monolithic ceramics.4.2 Velocity measurements described herein can assess subtle variations in porosity within a given material or component, as,for example, in ceramic superconductors and structural ceramic specimens (2, 3).4.3 In addition to ceramics and ceramic composites,

27、 the velocity measurements described herein may be applied topolycrystalline and single crystal metals, metal matrix composites, and polymer matrix composites.4.4 An alternative technique for velocity measurement is given in Practice E494.NOTE 1B1 and B2 are first and second back surface echoes, res

28、pectively, and T is time interval between the echoes.FIG. 1 Cross Section of Buffered Ultrasonic Probe (a) and Principle Echoes (b) for Velocity MeasurementC1331 1825. Personnel Qualifications5.1 It is recommended that nondestructive evaluation/examination personnel applying this test methodIf speci

29、fied in thecontractual agreement, personnel performing examinations to this practice shall be qualified in accordance with a nationally-recognized nationally or internationally recognized NDT personnel qualification practice or standard such as ASNTANSI/ASNT-CP-189, SNT-TC-1A, MIL STD 410, NAS 410,

30、ISO 9712, or a similar document. The qualification document and certified by theemployer or certifying agency, as applicable. The practice or standard used and its applicable revision(s) shouldrevision shall bespecifiedidentified in a contractual agreement.the contractual agreement between the using

31、 parties.5.2 Knowledge of the principles of ultrasonic testing is required. Personnel applying this test method practice should beexperienced practitioners of ultrasonic examinations and associated methods for signal acquisition, processing, and interpretation.5.3 Personnel should have proficiency i

32、n computer signal processing and the use of digital methods for time and frequencydomain signal analysis. Familiarity with Fourier and associated transforms for ultrasonic spectrum analysis is required.6. Qualification of Nondestructive Agencies6.1 If specified in the contractual agreement, NDT agen

33、cies shall be qualified and evaluated as described in Specification E543.The applicable edition of Specification E543 shall be specified in the contractual agreement.7. Apparatus and Test Sample7.1 Instrumentation (Fig. 1 and Fig. 2) for broadband cross-correlation pulse-echo ultrasonic velocity mea

34、surement shouldinclude the following:7.1.1 Buffered Probe:7.1.1.1 The buffer rod, which is an integral part of the probe (search unit), should be a right cylinder with smooth flat endsnormal to the axis of the probe.7.1.1.2 The center frequency of the buffered probe should produce a wavelength withi

35、n the sample that is less than one fifthof the thickness of the sample.7.1.1.3 The buffer rod length, that is, time delay should be three times the interval between two successive back surface echoes.7.1.1.4 The wave mode may be either longitudinal or shear.7.1.2 Pulser-Receiver, with a bandwidth th

36、at is at least twice that of the buffered probe. The bandwidth should includefrequencies in the range from 100 kHz to over 100 MHz.7.1.2.1 The pulser-receiver should have provisions for controlling the pulse repetition rate, pulse energy level, pulse damping,and received signal gain.7.1.2.2 The puls

37、er-receiver should provide a synchronization pulse and signal output connector.7.1.3 Waveform Digitizing Oscilloscope (A/D Board), bus programmable, to window and digitize the echo waveforms.7.1.3.1 Aminimum 512-element waveform array with a maximum data sampling interval of 1.95 ns is recommended.

38、For betterwaveform resolution, a 1024-element array with a data sampling interval of 0.97 ns may be needed.7.1.3.2 Vertical Amplifier, bus programmable module.7.1.3.3 Time Base, bus programmable module with a resolution of at least 5 ns per division and several time base rangesincluding a fundamenta

39、l time base of at least 200 ns.FIG. 2 Instrumentation Diagram for Acquiring and Separately Windowing Two Successive Back Surface Echoes, B1 and B2, for Cross-Correlation Velocity MeasurementC1331 1837.1.4 Digital Time Delay Module, bus programmable, to introduce a known time delay between the start

40、of two separate timegates, that is, windows each of which containing one of two successive back surface echoes.7.1.4.1 Separate windows are preferred for waveform digitization. Each waveform should occupy from 60 to 80 % of thewindow.7.1.4.2 The time synthesizer should have an accuracy of 61 ns with

41、 a precision of 60.1 ns.7.1.5 Video Monitors, (optional) one analog, one digital for real-time visual inspection of echo waveforms and for makinginteractive manual adjustments to the data acquisition controls.7.1.6 Computer, with adequate speed and storage capacity to provide needed software control

42、, data storage, and graphicscapability. The software should include a fast Fourier transform (FFT) algorithm package containing the cross-correlationalgorithm.7.1.7 Couplant Layer, to establish good signal transfer between the buffer rod and test sample. The layer should be as thin aspossible to min

43、imize couplant resonances and distortion of the echo waveforms.7.1.7.1 The couplant should not be absorbed by or be otherwise deleterious to the test sample.7.1.7.2 Dry coupling with a thin polymer may be used where liquid contamination by or absorption of liquids by the test sampleor part must be a

44、voided.7.2 The test sample or part should have flat parallel opposing surfaces in the region where the velocity measurements are made.This will assure good coupling between the transducer and sample and also produce valid echoes for velocity measurements.7.2.1 Lack of precision in the measurement of

45、 the test sample thickness can undermine the nanosecond precision with whichpulse-echo travel times can be measured.Therefore, the sample thickness should be measurable to an accuracy of60.1 % or better.7.2.2 For most engineering solids, the sample thickness should be at least 2.5 mm. There is a pra

46、ctical upper bound on samplethickness, for example, if the sample is too thick, there may be considerable signal attenuation, beam spreading, and dispersionthat render the signal useless.8. Procedure8.1 Use instrument control software routines to start and control the interface bus; perform procedur

47、es such as optimizingintensity, voltage, and time on the waveform digitizing oscilloscope; control the digital time delay module; and acquire, store, andprocess data.8.1.1 A cross-correlation algorithm should be part of the FFT software.8.1.2 The arguments needed to implement the cross-correlation a

48、lgorithm are the time domain waveform arrays, that is,digitized echoes B1 and B2 (Fig. 1).8.2 Prepare samples with front and back surfaces that are sufficiently smooth, flat, and parallel to allow measurement of the testsample thickness to an accuracy of 0.1 % or better.8.3 Couple the sample to the

49、transducer to obtain two strong back surface echoes.8.3.1 Apply pressure to minimize the couplant layer thickness. A backing fixture may be necessary to apply pressure.8.3.2 Care shall be taken to avoid coupling the sample to the backing fixture and thereby losing echo signal strength by leakage.8.3.3 A dry, hard rubber or composite material with a rough-machined or sawtooth surface is recommended for the backingfixture.NOTE 1Time delay, W, between the two window start times is predetermined. Time interval, T, between echoes B1 and B2 is calculated from T

展开阅读全文
相关资源
猜你喜欢
  • API TR 407-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Mysid Shrimp (Mysidopsis Bahia) Under Static Renewal Conditions TSCA Guideline Section Mark 797.1930《更换的静水三戊甲基醚对糠虾.pdf API TR 407-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Mysid Shrimp (Mysidopsis Bahia) Under Static Renewal Conditions TSCA Guideline Section Mark 797.1930《更换的静水三戊甲基醚对糠虾.pdf
  • API TR 407-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Mysid Shrimp (Mysidopsis Bahia) Under Static Renewal Conditions TSCA Guideline Section Mark 797.1930《更换的静水三戊甲基醚对糠虾_1.pdf API TR 407-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Mysid Shrimp (Mysidopsis Bahia) Under Static Renewal Conditions TSCA Guideline Section Mark 797.1930《更换的静水三戊甲基醚对糠虾_1.pdf
  • API TR 408-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Rainbow Trout (Oncorhynchus Mykiss) Under Flow-Through Conditions TSCA Guideline Section Mark 797.1400《在流水环境下三戊甲基醚.pdf API TR 408-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Rainbow Trout (Oncorhynchus Mykiss) Under Flow-Through Conditions TSCA Guideline Section Mark 797.1400《在流水环境下三戊甲基醚.pdf
  • API TR 408-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Rainbow Trout (Oncorhynchus Mykiss) Under Flow-Through Conditions TSCA Guideline Section Mark 797.1400《在流水环境下三戊甲基醚_1.pdf API TR 408-1995 Tert-Amyl Methyl Ether (TAME) - Acute Toxicity to Rainbow Trout (Oncorhynchus Mykiss) Under Flow-Through Conditions TSCA Guideline Section Mark 797.1400《在流水环境下三戊甲基醚_1.pdf
  • API TR 409-1995 Primary Skin Irritation Study In Rabbits of API 91-01 and PS-6 Unleaded Test Gasolines (CAIS Abstract No 42-2014)《兔子皮肤刺激性初级的研究 API 91-01 and PS-6无铅汽油(CAIS 摘要42-2014.pdf API TR 409-1995 Primary Skin Irritation Study In Rabbits of API 91-01 and PS-6 Unleaded Test Gasolines (CAIS Abstract No 42-2014)《兔子皮肤刺激性初级的研究 API 91-01 and PS-6无铅汽油(CAIS 摘要42-2014.pdf
  • API TR 410-1996 Chromosome Aberrations in Chinese Hamster Ovary (CHO) Cells Exposed to Tertiary Amyl Methyl Ether (TAME)《由于细胞在三戊甲基醚暴露过 中国仓鼠的卵巢染色体畸变》.pdf API TR 410-1996 Chromosome Aberrations in Chinese Hamster Ovary (CHO) Cells Exposed to Tertiary Amyl Methyl Ether (TAME)《由于细胞在三戊甲基醚暴露过 中国仓鼠的卵巢染色体畸变》.pdf
  • API TR 411-1996 Chinese Hamster Ovary (CHO) HGPRT Mutation Assay of Tertiary Amyl Methyl Ether (TAME)《中国仓鼠的卵巢 三戊甲基醚导致HGPRT基因突变分析》.pdf API TR 411-1996 Chinese Hamster Ovary (CHO) HGPRT Mutation Assay of Tertiary Amyl Methyl Ether (TAME)《中国仓鼠的卵巢 三戊甲基醚导致HGPRT基因突变分析》.pdf
  • API TR 412-1998 A Range-Finding Developmental Inhalation Toxicity Study of Unleaded Gasoline Vapor Condensate in Rats and Mice Via Whole-Body Exposers (Includes API TR 414)《通过把老鼠全身.pdf API TR 412-1998 A Range-Finding Developmental Inhalation Toxicity Study of Unleaded Gasoline Vapor Condensate in Rats and Mice Via Whole-Body Exposers (Includes API TR 414)《通过把老鼠全身.pdf
  • API TR 412-1998 A Range-Finding Developmental Inhalation Toxicity Study of Unleaded Gasoline Vapor Condensate in Rats and Mice Via Whole-Body Exposers (Includes API TR 414)《通过把老鼠全身_1.pdf API TR 412-1998 A Range-Finding Developmental Inhalation Toxicity Study of Unleaded Gasoline Vapor Condensate in Rats and Mice Via Whole-Body Exposers (Includes API TR 414)《通过把老鼠全身_1.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1