1、Designation: D6319 10Standard Specification forNitrile Examination Gloves for Medical Application1This standard is issued under the fixed designation D6319; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、 A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers certain requirements for nitrilerubber gloves used in conducting medical examinations anddiagnostic and th
3、erapeutic procedures.1.2 This specification covers nitrile rubber examinationgloves that fit either hand, paired gloves, and gloves by size. Italso provides for packaged sterile or nonsterile or bulk non-sterile nitrile rubber examination gloves.1.3 This specification is similar to that of Specifica
4、tionD3578 for rubber examination gloves.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.2. Referenced Documents2.1 ASTM Standards:2D412 Test Methods for Vulcanized Rubber and Thermo-plastic ElastomersTensionD573 Test Method
5、for RubberDeterioration in an AirOvenD3578 Specification for Rubber Examination GlovesD3767 Practice for RubberMeasurement of DimensionsD5151 Test Method for Detection of Holes in MedicalGlovesD6124 Test Method for Residual Powder on MedicalGloves2.2 ISO Standard:ISO 2859 Sampling Procedures and Tab
6、les for Inspectionby Attributes32.3 Other Documents:U.S. Pharmacopeia43. Significance and Use3.1 The specification is intended as a referee procedure forevaluating the performance and safety of nitrile rubber exami-nation gloves. The safe and proper use of nitrile rubberexamination gloves is beyond
7、the scope of this specification.4. Material4.1 Any nitrile rubber polymer compound may be used thatpermits the glove to meet the requirements of this specification.4.2 A lubricant that meets the current requirements of theU.S. Pharmacopeia for absorbable dusting powder may beapplied to the glove. Ot
8、her lubricants may be used if theirsafety and efficacy have been previously established.4.3 The inside and outside surface of the nitrile rubberexamination gloves shall be free of talc.5. Sampling5.1 For referee purposes, gloves shall be sampled fromfinished product, after sterilization when labeled
9、 sterile, andinspected in accordance with ISO 2859. The inspection levelsand acceptable quality levels (AQL) shall conform to thosespecified in Table 1, or as agreed upon between the purchaserand the seller, if the latter is more comprehensive.6. Performance Requirements6.1 Gloves, sampled in accord
10、ance with Section 5, shallmeet the following referee performance requirements:6.1.1 Product comply with requirements for sterility whentested in accordance with 7.2 when labeled sterile.6.1.2 Shall comply with freedom from holes when tested inaccordance with 7.3.6.1.3 Have consistent physical dimens
11、ions in accordancewith 7.4.6.1.4 Have acceptable physical property characteristics inaccordance with 7.5.6.1.5 Have a powder residue limit of 2.0 mg in accordancewith 7.66.1.6 Have a recommended maximum powder limit of 10mg/dm2in accordance with 7.7.1This specification is under the jurisdiction ofAS
12、TM Committee D11 on Rubberand is the direct responsibility of Subcommittee D11.40 on Consumer RubberProducts.Current edition approved May 1, 2010. Published June 2010. Originallyapproved in 1999. Last previous edition approved in 2005 as D6319 00a (2005)1.DOI: 10.1520/D6319-10.2For referenced ASTM s
13、tandards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute, 25 W. 43rd St., 4thFloor, Ne
14、w York, NY 10036.4U. S. Pharmacopeia, latest edition, Mack Publishing Co., Easton, PA 19175.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.7. Referee Test Methods7.1 The following tests shall be conducted to ensure therequirements o
15、f Section 6, as prescribed in Table 1:7.2 Sterility TestTesting for sterility shall be conducted inaccordance with the latest edition of the U.S. Pharmacopeia.7.3 Freedom from HolesTesting for freedom from holesshall be conducted in accordance with Test Method D5151.7.4 Physical Dimensions Test:7.4.
16、1 The gloves shall comply with the dimension require-ments prescribed in Table 2.7.4.2 The length shall be expressed in millimetres as mea-sured from the tip of the middle finger to the outside edge ofthe cuff.7.4.3 The width of the palm shall be expressed in millime-tres as measured at a level betw
17、een the base of the index fingerand the base of the thumb. Values of width per size other thanlisted shall meet the stated tolerance specified in Table 2.7.4.4 The minimum thickness shall be expressed in milli-metres as specified in Table 2 when using a dial or digitalmicrometer that meets requireme
18、nts described in Test MethodsD412 and Practice D3767, and in the locations indicated in Fig.1. For referee tests, cutting the glove is necessary to obtainsingle-thickness measurements. (See Practice D3767 for moreinformation.)7.5 Physical Requirements Test:7.5.1 Before and after accelerated aging, t
19、he gloves shallconform to the physical requirements specified in Table 3.Tests shall be conducted in accordance with Test MethodsD412. Die C is recommended.7.5.2 Accelerated AgingThe gloves shall be aged inaccordance with Test Method D573. Test the gloves in accor-dance with either one of the follow
20、ing methods:7.5.2.1 After being subjected to a temperature of 70 6 2Cfor 166 6 2 h, the tensile strength and ultimate elongation shallnot be less than the values specified in Table 3. This methodshall be the conditions for referee tests.7.5.2.2 After being subjected to a temperature of 100 6 2Cfor 2
21、2 6 0.3 h, the tensile strength and ultimate elongationshall not be less than the values specified in Table 3.7.6 Powder Free GlovesDetermine the powder residueusing Test Method D6124.7.7 Powdered Gloves:7.7.1 Determine the recommended maximum powder limitusing Test Method D6124 for powdered gloves.
22、7.7.2 Determine the square decimetres for the glove size asin the paragraph on determining the square decimetres of glovesize in Specification D3578.8. Acceptance8.1 Gloves will be considered to meet the referee perfor-mance requirements when test results conform to the require-ments prescribed in T
23、able 1.8.2 Retests or reinspections are permissible under the pro-vision of the U.S. Pharmacopeia and ISO 2859.9. Packaging and Package Marking9.1 Sterile Packaging:9.1.1 The unit of packaging shall normally be one glove orone pair of gloves.9.1.2 A glove or pair of gloves, normally, shall be enclos
24、edin an inner wallet or wrapper. The wrapper shall be of sufficientsize when opened to provide a field for glove-donning pur-poses.9.1.3 The glove or pair of gloves, and accompanyingwrapper if utilized, shall be totally enclosed in an outerpackage that will allow sterilization of the product.9.1.4 T
25、he outer package shall have a method of closuresufficient to ensure the sterility of the product until opened ordamaged.9.1.5 The outer package shall have sufficient strength andintegrity to withstand normal transportation and storage withinthe intermediate or shipping cartons, or both.9.1.6 The met
26、hod of closure of the outer package shall besuch that prior opening will be detectable by the user.TABLE 1 Performance RequirementsCharacteristic Related Defects Inspection Level AQLSterility fails sterilityAN/AFreedom from holes holes G-1 2.5Dimensions width, length, and thickness S-2 4.0Physical p
27、roperties before aging, afteraccelerated agingS-2 4.0Powder-free Residue exceeds maximum limit N=5 N/APowder Amount exceeds recommendedmaximum limitN=2 N/AASee U.S. Pharmacopeia.TABLE 2 Dimensions and TolerancesNOTESizing that falls within the tolerance overlaps between two sizes may be labeled as a
28、 size range including both sizes, for example, small/mediumand medium/large.DesignationSizeTolerance, mm6612 7712 8812 9Width by size 75 83 89 95 102 108 114 66Width by x-small small Unisize medium large X-large 61070 80 85 95 110 120Length 220 220 230 230 230 230 minThickness, mm:finger 0.05 minpal
29、m 0.05 minD6319 1029.1.7 None of the packaging material shall contain anymaterial likely to impair the quality and use of the gloves.9.1.8 Intermediate cartons and shipping cases shall be ofsufficient strength to maintain the quality and sterility of theproduct during normal transportation and stora
30、ge.9.2 Nonsterile and Bulk Packaging:9.2.1 The gloves shall be enclosed in an outer package thathas sufficient strength to withstand normal transportation andstorage within the cartons or shipping cases, or both.9.2.2 None of the packaging material shall contain anymaterial likely to impair the qual
31、ity and use of the gloves.9.2.3 Cartons and shipping cases shall be of sufficientstrength to maintain the quality of the product during normaltransportation and storage.9.3 Package Marking:9.3.1 Sterile packages shall bear markings for the contentsto include the glove size, instructions for opening,
32、 the legend“sterile,” and a manufacturing lot number.9.3.2 Nonsterile and bulk packages shall bear markings forthe contents to include the glove size and a manufacturing lotnumber.9.3.3 The outermost case shall be labeled with the glove sizeand a manufacturing lot number. Sterile product cases shall
33、also be marked with the legend “sterile.”9.3.4 All levels of packaging shall conform to all appropri-ate government labeling regulations.10. Keywords10.1 examination gloves; nitrile; rubberFIG. 1 Location of Thickness MeasurementsTABLE 3 Physical RequirementsBefore Aging After Accelerated AgingTensi
34、le Strength Ultimate Elongation Tensile Strength Ultimate Elongation14 MPa, min 500 % min 14 MPa min 400 % minD6319 103ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressl
35、y advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, e
36、ither reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may atte
37、nd. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. I
38、ndividual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/COPYRIGHT/).D6319 104