ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf

上传人:ideacase155 文档编号:518621 上传时间:2018-12-03 格式:PDF 页数:5 大小:93.68KB
下载 相关 举报
ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf_第1页
第1页 / 共5页
ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf_第2页
第2页 / 共5页
ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf_第3页
第3页 / 共5页
ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf_第4页
第4页 / 共5页
ASTM D5002-1999(2010) Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer《数字密度分析仪用原油密度和相对密度的标准试验方法》.pdf_第5页
第5页 / 共5页
亲,该文档总共5页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: D5002 99 (Reapproved 2010)Standard Test Method forDensity and Relative Density of Crude Oils by DigitalDensity Analyzer1This standard is issued under the fixed designation D5002; the number immediately following the designation indicates the year oforiginal adoption or, in the case of r

2、evision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the densityor relative density of crude oils that ca

3、n be handled in a normalfashion as liquids at test temperatures between 15 and 35C.This test method applies to crude oils with high vapor pressuresprovided appropriate precautions are taken to prevent vaporloss during transfer of the sample to the density analyzer.1.2 This test method was evaluated

4、in round robin testingusing crude oils in the 0.75 to 0.95 g/mL range. Lighter crudeoil can require special handling to prevent vapor losses.Heavier crudes can require measurements at higher tempera-tures to eliminate air bubbles in the sample.1.3 The values stated in SI units are to be regarded ass

5、tandard. No other units of measurement are included in thisstandard. The accepted units of measurement of density aregrams per millilitre and kilograms per cubic metre.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of

6、 the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific warningstatements are given in 7.4, 7.5, and 7.6.2. Referenced Documents2.1 ASTM Standards:2D941 Test Method for Density and Relative Den

7、sity (Spe-cific Gravity) of Liquids by Lipkin Bicapillary Pycnom-eter3D1193 Specification for Reagent WaterD1217 Test Method for Density and Relative Density (Spe-cific Gravity) of Liquids by Bingham PycnometerD1250 Guide for Use of the Petroleum Measurement TablesD4052 Test Method for Density, Rela

8、tive Density, and APIGravity of Liquids by Digital Density MeterD4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4177 Practice for Automatic Sampling of Petroleum andPetroleum ProductsD4377 Test Method for Water in Crude Oils by Potentio-metric Karl Fischer Titration3. Terminolo

9、gy3.1 Definitions:3.1.1 densitymass per unit volume at a specified tempera-ture.3.1.2 relative densitythe ratio of the density of a materialat a stated temperature to the density of water at a statedtemperature.4. Summary of Test Method4.1 Approximately 0.7 mL of crude oil sample is introducedinto a

10、n oscillating sample tube and the change in oscillatingfrequency caused by the change in the mass of the tube is usedin conjunction with calibration data to determine the density ofthe sample.5. Significance and Use5.1 Density is a fundamental physical property that can beused in conjunction with ot

11、her properties to characterize thequality of crude oils.5.2 The density or relative density of crude oils is used forthe conversion of measured volumes to volumes at the standardtemperatures of 15C or 60F and for the conversion of crudemass measurements into volume units.5.3 The application of the d

12、ensity result obtained from thistest method, for fiscal or custody transfer accounting calcula-tions, can require measurements of the water and sedimentcontents obtained on similar specimens of the crude oil parcel.6. Apparatus6.1 Digital Density AnalyzerA digital analyzer consistingof a U-shaped, o

13、scillating sample tube and a system forelectronic excitation, frequency counting, and display. Theanalyzer must accommodate the accurate measurement of the1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of Subcomm

14、itteeD02.04.0D on Physical and Chemical Methods.Current edition approved Sept. 15, 2010. Published November 2010. Originallyapproved in 1989. Last previous edition approved in 2005 as D500299(2005).DOI: 10.1520/D5002-99R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orconta

15、ct ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn. The last approved version of this historical standard is referencedon www.astm.org.1Copyright ASTM International, 100 Barr Har

16、bor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.sample temperature during measurement or must control thesample temperature as described in 6.2 and 6.5. The instrumentshall be capable of meeting the precision requirements de-scribed in Test Method D4052.6.2 Circulating Consta

17、nt-Temperature Bath, capable ofmaintaining the temperature of the circulating liquid constantto 60.05C in the desired range. Temperature control can bemaintained as part of the density analyzer instrument package.6.3 Syringes, at least 2 mL in volume with a tip or anadapter tip that will fit the inl

18、et of the density analyzer.6.4 Flow-Through or Pressure Adapter, for use as analternative means of introducing the sample into the densitymeter.6.5 Thermometer, calibrated and graduated to 0.1C, and athermometer holder that can be attached to the instrument forsetting and observing the test temperat

19、ure. In calibrating thethermometer, the ice point and bore corrections should beestimated to the nearest 0.05C. Precise setting and control ofthe test temperature in the sample tube is extremely important.An error of 0.1C can result in a change in density of one in thefourth significant figure.7. Re

20、agents and Materials7.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are availabl

21、e.4Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Ty

22、pe II of Specification D1193.7.3 Water, redistilled, freshly boiled and cooled reagentwater for use as a primary calibration standard.7.4 Acetone, for flushing and drying the sample tube.(WarningExtremely flammable.)7.5 Petroleum Naphtha, for flushing viscous petroleumsamples from the sample tube. (

23、WarningExtremely flam-mable.)NOTE 1Suitable solvent naphthas are marketed under various desig-nations such as “petroleum ether,” “ligroine,” or “precipitation naphtha.”7.6 n-Nonane, n-tridecane or cyclohexane, 99 % purity orbetter, or similar pure material for which the density is knownprecisely fro

24、m literature references or by direct determinationin accordance with Test Method D941 or D1217.(WarningExtremely flammable.)8. Sampling, Test Specimens, and Test Units8.1 Sampling is defined as all the steps required to obtain analiquot of the contents of any pipe, tank or other system, and toplace

25、the sample into the laboratory test container. Thelaboratory test container and sample volume shall be ofsufficient dimensions to allow mixing as described in 8.3.1.Mixing is required to obtain a homogeneous sample foranalysis.8.2 Laboratory SampleUse only representative samplesobtained as specified

26、 in Practices D4057 or D4177 for this testmethod.8.3 Test SpecimenThe aliquot of sample obtained fromthe laboratory sample and delivered to the density analyzersample tube. The test specimen is obtained as follows:8.3.1 Mix the sample of crude oil to homogenize anysediment and water present. The mix

27、ing may be accomplishedas described in Practice D4177 or Test Method D4377. Mixingat room temperature in an open container can result in the lossof light ends, so mixing in closed, pressurized containers or atsub-ambient temperatures is recommended.8.3.2 Draw the test specimen from a properly mixed

28、labo-ratory sample using an appropriate syringe.Alternatively, if theproper density analyzer attachments and connecting tubes areused then the test specimen can be delivered directly to theanalyzers sample tube from the mixing container.9. Preparation of Apparatus9.1 Set up the density analyzer and

29、constant temperaturebath following the manufacturers instructions. Adjust the bathor internal temperature control so that the desired test tempera-ture is established and maintained in the sample compartmentof the analyzer. Calibrate the instrument at the same tempera-ture at which the density of th

30、e sample is to be measured.10. Calibration of Apparatus10.1 Calibrate the instrument when first setting up andwhenever the test temperature is changed. Thereafter, conductcalibration checks at least weekly during routine operation ormore frequently as may be dictated by the nature of the crudeoils b

31、eing measured (see 10.3).10.2 Initial calibration, or calibration after a change in testtemperature, necessitates calculation of the values of theConstants A and B from the periods of oscillation, (T),observed when the sample cell contains certified referenceliquids such as air and double-distilled

32、boiled water. Othercalibrating materials such as n-nonane, n-tridecane, cyclohex-ane, and n-hexadecane (for high temperature applications) canalso be used as appropriate.10.2.1 While monitoring the oscillator period, T, flush thesample tube with petroleum naphtha, followed with an acetoneflush and d

33、ry with dry air. Continue drying until the displayexhibits a steady reading. In cases where saline componentscan be deposited in the cell, flush with distilled water followedby acetone and dry air. Contaminated or humid air can affectthe calibration. When these conditions exist in the laboratory,pas

34、s the air used for calibration through a suitable purificationand drying train. In addition, the inlet and outlet ports for theU-tube must be plugged during measurement of the calibrationair to prevent ingress of moist air.10.2.2 Allow the dry air in the U-tube to come to thermalequilibrium with the

35、 test temperature and record the T-valuefor air.4Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, DC. For Suggestions on the testing of reagents notlisted by the American Chemical Society, see Annual Standards for LaboratoryChemicals, BDH Ltd., Poole

36、, Dorset, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.D5002 99 (2010)210.2.3 Introduce about 0.7 mL of freshly boiled and cooleddouble-distilled water into the sample tube from the bottomopening using a suitable syringe. Th

37、e water must be free ofeven the smallest air or gas bubbles. The sample tube shall becompletely full. Allow the water to reach thermal equilibriumat the test temperature and record the T-value for water and thetest temperature.10.2.4 Alternatively introduce one of the hydrocarbon cali-bration standa

38、rds and measure the T-value as in 10.2.3.10.2.5 Calculate the density of air at the temperature of testusing the following equation:da5 0.001293273/TP/760 g/mL (1)where:T = temperature, K, andP = barometric pressure, torr.10.2.6 Determine the density of water at the temperature oftest by reference t

39、o Table 1.10.2.7 Alternatively record the density at the test tempera-ture for the hydrocarbon calibrant used in 10.2.4 as obtainedfrom an appropriate reference source or from direct determi-nation (see 7.6).10.2.8 Using the observed T-values and the reference valuesfor water and air, calculate the

40、values of the Constants A and Busing the following equations:A 5 Tw22 Ta2#/dw2 da# (2)B 5 Ta22 A 3 da! (3)where:Tw= observed period of oscillation for cell containingwater,Ta= observed period of oscillation for cell containing air,dw= density of water at test temperature, andda= density of air at te

41、st temperature.Alternatively, use the T and d values for the other referenceliquid if one is used.10.2.9 If the instrument is equipped to calculate densityfrom the Constants A and B and the observed T-value from thesample, then enter the constants in the instrument memory inaccordance with the manuf

42、acturers instructions.10.2.10 Check the calibration and adjust if needed byperforming the routine calibration check described in 10.3.10.2.11 To calibrate the instrument to determine relativedensity, that is, the density of the sample at a given temperaturereferred to the density of water at the sam

43、e temperature, follow10.2.1-10.2.9, but substitute 1.000 for dwin performing thecalculations described in 10.2.8.10.3 Since some crude oils can be difficult to remove fromthe sample tube, frequent calibration checks are recommended.These checks and any subsequent adjustments to Constants Aand B can

44、be made if required, without repeating the calcula-tion procedure.NOTE 2The need for a change in calibration is generally attributableto deposits in the sample tube that are not removed by the routine flushingprocedure.Although this condition can be compensated for by adjustingAand B, as described b

45、elow, it is good practice to clean the tube with warmchromic acid solution (WarningCauses severe burns. A recognizedcarcinogen.) whenever a major adjustment is required. Chromic acidsolution is the most effective cleaning agent; however, surfactant-typecleaning fluids have also been used successfull

46、y.10.3.1 Flush and dry the sample tube as described in 10.2.1and allow the display to reach a steady reading. If the displaydoes not exhibit the correct T-value or density for air at thetemperature of test, repeat the cleaning procedure or adjust thevalue of Constant B commencing with the last decim

47、al placeuntil the correct density is displayed.10.3.2 If adjustment to Constant B was necessary in 10.3.1then continue the recalibration by introducing freshly boiledand cooled double-distilled water into the sample tube asdescribed in 10.2.3 and allowing the display to reach a steadyreading. If the

48、 instrument has been calibrated to display thedensity, adjust the reading to the correct value for water at thetest temperature (see Table 1) by changing the value ofConstant A, commencing with the last decimal place. If theinstrument has been calibrated to display the relative density,adjust the re

49、ading to the value 1.0000.NOTE 3In applying this periodic calibration procedure, it has beenfound that more than one value each for A and B, differing in the fourthdecimal place, will yield the correct reading for the density of air andwater. The setting chosen would then be dependent upon whether it wasapproached from a higher or lower value. The setting selected by thismethod could have the effect of altering the fourth place of the readingobtained for a sample.10.4 Some analyzer models are designed to display themeasured period of oscillation only (T

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1